Page 13 highlights
• Pressure sensors can be damaged and/or personnel injury can occur due to improper application of pressure. Vacuum should not be applied to any gauge pressure sensor. The Product display shows "OL" when an inappropriate pressure is applied. If "OL" is shown on any pressure display, the pressure should be reduced or vented immediately to prevent Product damage or possible personnel injury. "OL" is shown when the pressure exceeds 110 % of the nominal range of the sensor or when a vacuum in excess of 2 PSI is applied on gauge range sensors. • Push the ZERO button to zero the pressure sensor when vented to atmospheric pressure. Pressure Calibrator Safety Information • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures that exceed the battery manufacturer's specification. If the batteries are not removed, battery leakage can damage the Product. • Replace the batteries when the low battery indicator shows to prevent incorrect measurements. • Be sure that the battery polarity is correct to prevent battery leakage. • Repair the Product before use if the battery leaks. • The battery door must be closed and locked before you operate the Product. • Use only specified replacement parts. 3 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664