Fluke 719PRO-300G Product Manual - Page 13

Pressure Calibrator

Page 13 highlights

• Pressure sensors can be damaged and/or personnel injury can occur due to improper application of pressure. Vacuum should not be applied to any gauge pressure sensor. The Product display shows "OL" when an inappropriate pressure is applied. If "OL" is shown on any pressure display, the pressure should be reduced or vented immediately to prevent Product damage or possible personnel injury. "OL" is shown when the pressure exceeds 110 % of the nominal range of the sensor or when a vacuum in excess of 2 PSI is applied on gauge range sensors. • Push the ZERO button to zero the pressure sensor when vented to atmospheric pressure. Pressure Calibrator Safety Information • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures that exceed the battery manufacturer's specification. If the batteries are not removed, battery leakage can damage the Product. • Replace the batteries when the low battery indicator shows to prevent incorrect measurements. • Be sure that the battery polarity is correct to prevent battery leakage. • Repair the Product before use if the battery leaks. • The battery door must be closed and locked before you operate the Product. • Use only specified replacement parts. 3 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Pressure Calibrator
Safety Information
3
Pressure sensors can be damaged
and/or personnel injury can occur due to
improper application of pressure.
Vacuum should not be applied to any
gauge pressure sensor. The Product
display shows “OL” when an
inappropriate pressure is applied. If “OL”
is shown on any pressure display, the
pressure should be reduced or vented
immediately to prevent Product damage
or possible personnel injury. “OL” is
shown when the pressure exceeds 110 %
of the nominal range of the sensor or
when a vacuum in excess of 2 PSI is
applied on gauge range sensors.
Push the ZERO button to zero the
pressure sensor when vented to
atmospheric pressure.
Remove the batteries if the Product is
not used for an extended period of time,
or if stored in temperatures that exceed
the battery manufacturer’s specification.
If the batteries are not removed, battery
leakage can damage the Product.
Replace the batteries when the low
battery indicator shows to prevent
incorrect measurements.
Be sure that the battery polarity is
correct to prevent battery leakage.
Repair the Product before use if the
battery leaks.
The battery door must be closed and
locked before you operate the Product.
Use only specified replacement parts.
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com