Fluke 719PRO-300G Product Manual - Page 62
Cprt_r0, Cprt_r0?, Damp?, Display, Display
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 62 highlights
719PRO Users Manual CPRT_R0 Sets the 0 ° resistance, R0, in the custom RTD. The value must be entered with a unit label. Refer to Table for assistance. For example: CPRT_R0 100 OHM sets R0 to 100 Ω. CPRT_R0? Returns the value for the resistance in custom RTD. The above example returns: 1.000000E+02, OHM DAMP Turns on or off the damping function. For example: If DAMP ON is sent, this will turn on the damping function. DAMP? Returns the current state of the damping function. For example: If DAMP? is sent, it will return ON if the damping function is on. DISPLAY Turns on or off the indicated display. For example: If DISPLAY LOWER is sent ON, this will turn on the lower display. DISPLAY? Returns the current state of the each of the displays. For example: If DISPLAY? is sent, it will return ON, ON, ON if the all the displays are on. 52 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664