Fluke 719PRO-300G Product Manual - Page 39
Transmitter Calibration, mA Input Function
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 39 highlights
Transmitter Calibration mA Input Function The mA input function reads back the 4 mA to 20 mA output from the device being calibrated. This can be done in one of two ways: 1. Passively - Where the device under test directly regulates or generates 4 mA to 20 mA and can be read by the Product. Pressure Calibrator Transmitter Calibration 2. Actively - Where the Product supplies 24 V dc loop power to the device under test to power the device while reading the resulting 4 mA to 20 mA signal. hix048.eps hix047.eps 29 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
Pressure Calibrator
Transmitter Calibration
29
Transmitter Calibration
mA Input Function
The mA input function reads back the 4 mA to 20 mA
output from the device being calibrated. This can be done
in one of two ways:
1.
Passively – Where the device under test directly
regulates or generates 4 mA to 20 mA and can be
read by the Product.
hix047.eps
2.
Actively – Where the Product supplies 24 V dc loop
power to the device under test to power the device
while reading the resulting 4 mA to 20 mA signal.
hix048.eps