Fluke 719PRO-300G Product Manual - Page 18
Display
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 18 highlights
719PRO Users Manual Notes When is pushed to turn on the Product, a short startup self-check routine is run. During that routine, the display shows the current firmware revision level, auto-shutdown status, and the range of the internal pressure sensor. A maximum of 5 minutes warm-up is necessary for the product to reach rated accuracy. A longer warm-up period can be necessary for large changes in ambient temperature. See the "Zero Function Use" section for more about zeroing the pressure sensor displays. It is recommended that pressure ranges be zeroed each time the Product is started. Display The display has two main regions: • The menu bar (located at the bottom of the screen) is used to access the Product menu. • The main display has a maximum of three process measurement sub-regions. These sub-regions will be referred to as the UPPER, MIDDLE and LOWER displays. Figure 2 shows the location of the different display fields while Table 3 describes them. 3 1 4 2 Figure 2. Display hix007.eps 8 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664