Fluke 719PRO-300G Product Manual - Page 51

Set up the RS-232 Port for Remote Control, Remote Operation

Page 51 highlights

Set up the RS-232 Port for Remote Control Note The RS-232 connection cable should not exceed 15 m unless the load capacitance measured at connection points is less than 2500 pF. Serial parameter values: • 9600 baud • 8 data bits • 1 stop bit • no parity • Xon/Xoff • EOL (End of Line) character or CR (Carriage Return) or both Note To use the remote control option, a custom miniature circular to USB interface cable, PN 4401616, must be purchased separately. Connect the Product to the computer, attach the miniature circular connector end of the cable to the pressure module port on the right side of the Product and the USB connector to the computer. The Product should be turned off prior to making the connection and then turned on. Pressure Calibrator Remote Operation For remote operation of the Product, connect it to a COM port on the PC as in Figure 11. Use a terminal emulator program and follow the subsequent procedure: 1. Start the terminal emulator program. 2. Select New Connection. 3. For Name enter "Fluke 719PRO". Select the serial port that the Product is connected to. 4. Enter the above information for port settings. 5. Select ASCII setup from File/Properties/Settings and mark these choices: • Echo typed characters locally • Wrap lines that exceed terminal width 6. Select Ok. 7. To see if the port works enter *IDN?. This command will return information on the Product. 41 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Pressure Calibrator
Remote Operation
41
Set up the RS-232 Port for Remote Control
Note
The RS-232 connection cable should not exceed
15 m unless the load capacitance measured at
connection points is less than 2500 pF.
Serial parameter values:
9600 baud
8 data bits
1 stop bit
no parity
Xon/Xoff
EOL (End of Line) character or CR (Carriage Return)
or both
Note
To use the remote control option, a custom
miniature circular to USB interface cable, PN
4401616, must be purchased separately.
Connect the Product to the computer, attach the
miniature circular connector end of the cable to the
pressure module port on the right side of the Product and
the USB connector to the computer. The Product should
be turned off prior to making the connection and then
turned on.
For remote operation of the Product, connect it to a COM
port on the PC as in Figure 11. Use a terminal emulator
program and follow the subsequent procedure:
1.
Start the terminal emulator program.
2.
Select New Connection.
3.
For Name enter “Fluke 719PRO”. Select the serial
port that the Product is connected to.
4.
Enter the above information for port settings.
5.
Select ASCII setup from File/Properties/Settings and
mark these choices:
Echo typed characters locally
Wrap lines that exceed terminal width
6.
Select Ok.
7.
To see if the port works enter *IDN?. This command
will return information on the Product.