Fluke 719PRO-300G Product Manual - Page 27
Activate and Deactivate a Display, Set the RTD Probe Type
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
Activate and Deactivate a Display From the Display Main Menu, push to access the Display Activation Menu. Pressure Calibrator Product Features Set the RTD Probe Type Push from the Probe Type Main Menu to access the RTD Probe selections. hix033.eps The {display} option can be used to select which display is active. The selected display and current on/off state are shown in the lower display as seen below. hix034.eps Push to save the changes and go back to the Main Menu. When a display is deactivated, its configuration is saved. When the display is activated, its configuration is compared against the configurations of the other currently-active displays. If the configurations are in conflict, the recalled display configuration is changed to prevent conflict. If all three displays are deactivated, the LOWER display comes on automatically. hix035.eps There are four probe types to select from: • P100-385 (select this type for use with Fluke- 720RTD probe accessory) • P100-392 • P100-JIS • CUSTOM Push to select the necessary probe type (see the figure below). Push to store the change and go to the Main Menu. Note The default probe type is PT100-385. hix036.eps 17 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664