From the Main Menu, push, to start the %ERROR option.
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 43 highlights
Example: A pressure transmitter under test is 30 psi (2 Bar) FullScale and regulates a corresponding 4 mA to 20 mA signal. Program a 0 psi to 30 psi pressure span into the Product and the Product calculates and shows the deviation or %Error from the expected 4 mA to 20 mA output. This eliminates the need for manual calculations and helps when it is difficult to set an exact pressure with an external pump. To use the %ERROR function: 1. From the Main Menu, push . 2. Push . 3. Push to start the %ERROR option. 4. Push to configure the option. The first option sets the Port. 5. Push to scroll through the port choices. 6. When finished, push . Pressure Calibrator Transmitter Calibration hix049.eps 7. Loop Power can be toggled on/off as seen below. Push when done. hix050.eps 33 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664