Fluke 719PRO-300G Product Manual - Page 21
Main Menu Functionality, Set the Active Display, Set Active Display Parameters
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
Main Menu Functionality There are three options on the Main Menu, shown below: hix021.eps Set the Active Display The active display is indicated by the center option on the Main Menu. Push from this menu to toggle the active display (UPPER, MIDDLE, or LOWER). Set Active Display Parameters To set the parameters of the active display, push from the Main Menu to go to the Active Display Configuration menu, shown below: Pressure Calibrator Product Features Push to change to the NEXT active display parameter. Only RTD and Pressure modes have a second parameter. RTDs can be read in Celsius or Fahrenheit. Pressures can be read in 15 engineering units. With a single display, the modes below are available. Push to access each mode: • [P1] = Pressure internal sensor • [EXT] = Pressure with external pressure module • [P1] ST = Switch Test with left side sensor • [EXT] ST = Switch Test with external pressure module hix022.eps Push to toggle through the selections for each parameter of the active display. The first parameter is Mode. Since voltage, current, and switch test modes all use the same jacks, two of these functions cannot be used at the same time. The ability to select some functions is limited based on what is already selected on a different display. 11 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664