Fluke 719PRO-300G Product Manual - Page 23

Zero Function Use, Internal Sensor and Pressure Module (non-absolute), Absolute Pressure, absolute

Page 23 highlights

Zero Function Use If a pressure mode is selected and pressure is within the zero limit, the Product zeros the active display when Z is pushed. The zero limits are within 10 % of the full-scale range of the selected sensor. If the display shows "OL," the zero function will not operate. Note Z is only used for pressure. Internal Sensor and Pressure Module (nonabsolute) When a sensor or module is selected on the active display, and Z is pushed, the Product subtracts the current reading from the output. The zero limits are within 10 % of the full-scale range of the selected sensor. If the display shows "OL," the zero function will not operate. Pressure Calibrator Product Features Absolute Pressure When an absolute pressure range is selected on the active display, and Z is pushed, the Product prompts for the barometric reference pressure as in the screen below. hix023.eps Push  and  to use the on-screen arrows to record the correct barometric reference pressure. The sensor port should be open (vented) to atmosphere while this procedure is done. 13 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Pressure Calibrator
Product Features
13
Zero Function Use
If a pressure mode is selected and pressure is within the
zero limit, the Product zeros the active display when
Z
is
pushed. The zero limits are within 10 % of the full-scale
range of the selected sensor. If the display shows “OL,”
the zero function will not operate.
Note
Z
is only used for pressure.
Internal Sensor and Pressure Module (non-
absolute)
When a sensor or module is selected on the active
display, and
Z
is pushed, the Product subtracts the
current reading from the output. The zero limits are within
10 % of the full-scale range of the selected sensor. If the
display shows “OL,” the zero function will not operate.
Absolute Pressure
When an absolute pressure range is selected on the
active display, and
Z
is pushed, the Product prompts for
the barometric reference pressure as in the screen below.
hix023.eps
Push
and
to use the on-screen arrows to record the
correct barometric reference pressure. The sensor port
should be open (vented) to atmosphere while this
procedure is done.