Fluke 719PRO-300G Product Manual - Page 23
Zero Function Use, Internal Sensor and Pressure Module (non-absolute), Absolute Pressure, absolute
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 23 highlights
Zero Function Use If a pressure mode is selected and pressure is within the zero limit, the Product zeros the active display when Z is pushed. The zero limits are within 10 % of the full-scale range of the selected sensor. If the display shows "OL," the zero function will not operate. Note Z is only used for pressure. Internal Sensor and Pressure Module (nonabsolute) When a sensor or module is selected on the active display, and Z is pushed, the Product subtracts the current reading from the output. The zero limits are within 10 % of the full-scale range of the selected sensor. If the display shows "OL," the zero function will not operate. Pressure Calibrator Product Features Absolute Pressure When an absolute pressure range is selected on the active display, and Z is pushed, the Product prompts for the barometric reference pressure as in the screen below. hix023.eps Push and to use the on-screen arrows to record the correct barometric reference pressure. The sensor port should be open (vented) to atmosphere while this procedure is done. 13 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664