Fluke 719PRO-300G Product Manual - Page 77

Specifications, Pressure Measurement, Range, Resolution, Accuracy, Comment, Function, Accuracy 1 year

Page 77 highlights

Specifications Pressure Measurement 719Pro-30G 719Pro-150G 719Pro-300G Temperature effect (all ranges) Function mA dc (simultation and measurement) Range -12 to 30 PSI/-0.8 to 2 bar -12 to 150 PSI/-0.8 to 10 bar -12 to 300 PSI/-0.8 to 20 bar Range 0 mA dc to 24 mA dc Volts dc (measurement only) 30 V dc Resolution 0.001 PSI 0.0001 bar 0.01 PSI 0.001 bar 0.01 PSI 0.001 bar Resolution 0.001 mA 0.001 V Temperature (measurement only, Pt100 RTD) Loop power supply Temperature Effect (all functions) -50 °C to 150 °C (-58 °F to 302 °F) 0.01 °C 0.01 °F 24 V dc N/A Pressure Calibrator Specifications Accuracy Comment ±0.025 % of full scale, 6 months ±0.035 % of full scale, 1 year Dry air and non-corrosive gas only Add ±0.002 % of full scale for temps outside of 15 °C to 35 °C Accuracy (1 year) ±0.015 % of reading ±2 counts Temperature ±0.1 °C (0.2 °F) 24 mA at 24 V Add ±0.002 % full scale/°C for temperatures outside of 15 °C to 35 °C No effect on accuracy on all functions from 15 °C to 35 °C Comment 1000 Ω maximum load in mA source, 26 V dc maximum voltage in mA simulate No ac voltage measurement, do not exceed 30 V dc ±0.25 °C (±0.45 °F) combined uncertainty when using 720 RTD probe (optional accessory) No effect on accuracy on all functions from 15 °C to 35 °C 67 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Pressure Calibrator
Specifications
67
Specifications
Pressure Measurement
Range
Resolution
Accuracy
Comment
719Pro-30G
-12 to 30 PSI/-0.8 to
2 bar
0.001 PSI
0.0001 bar
±
0.025 % of full scale,
6 months
±
0.035 % of full scale,
1 year
Dry air and non-corrosive
gas only
719Pro-150G
-12 to 150 PSI/-0.8 to
10 bar
0.01 PSI
0.001 bar
719Pro-300G
-12 to 300 PSI/-0.8 to
20 bar
0.01 PSI
0.001 bar
Temperature effect (all
ranges)
Add
±
0.002 % of full
scale for temps outside
of 15
°
C to 35
°
C
No effect on accuracy on
all functions from 15
°
C
to 35
°
C
Function
Range
Resolution
Accuracy (1 year)
Comment
mA dc
(simultation and
measurement)
0 mA dc to 24 mA dc
0.001 mA
±
0.015 % of reading
±
2
counts
1000
maximum load in
mA source, 26 V dc
maximum voltage in mA
simulate
Volts dc
(measurement only)
30 V dc
0.001 V
No ac voltage
measurement, do not
exceed 30 V dc
Temperature
(measurement only,
Pt100 RTD)
-50
°
C to 150
°
C (-58
°
F
to 302
°
F)
0.01
°
C
0.01
°
F
Temperature
±
0.1
°
C
(0.2
°
F)
±
0.25
°
C (
±
0.45
°
F)
combined uncertainty
when using 720 RTD
probe (optional
accessory)
Loop power supply
24 V dc
N/A
24 mA at 24 V
Temperature Effect (all
functions)
Add
±
0.002 % full
scale/
°
C for temperatures
outside of 15
°
C to 35
°
C
No effect on accuracy on
all functions from 15
°
C
to 35
°
C