Fluke 719PRO-300G Product Manual - Page 61
Calibrator Commands
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 61 highlights
Calibrator Commands CPRT_COEFA Use this command to entere a custom RTD into the Product. The numeric value entered after the command is set as the first coefficient of the polynomial used by the custom RTD. For example: CPRT_COEFA 3.908300E-03 enters 3.908300e-3 as coefficient A. CPRT_COEFA? Returns the number that was entered for the first coefficient of the polynomial used in the custom RTD. With the example above, CPRT_COEFA? returns: 3.908300E-03 CPRT_COEFB This command is used to enter a custom RTD into the Product. The numeric value entered after the command will be set as the second coefficient of the polynomial used by the custom RTD. For example: CPRT_COEFB -5.774999E-07 enters -5.774999E-07 as coefficient B. Pressure Calibrator Remote Operation CPRT_COEFB? Returns the number which was entered for the second coefficient of the polynomial used in the custom RTD. With the example above, CPRT_COEFB? returns: -5.774999E-07 CPRT_COEFC This command is used to enter a custom RTD into the Product. The numeric value entered after the command will be set as the first coefficient of the polynomial used by the custom RTD. For example: CPRT_COEFC -4.183000E-12 enters -4.183000E-12 as coefficient C. CPRT_COEFC? Returns the number that was entered for the third coefficient of the polynomial used in the custom RTD. The example above CPRT_COEFC? returns: -4.183000E-12 51 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664