Fluke 719PRO-300G Product Manual - Page 37
done slowly to ensure accurate readings., changes. Even with this enhanced sample rate
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 37 highlights
Note In switch-test mode the display update rate is increased to help capture pressure input changes. Even with this enhanced sample rate, pressurization of the device under test should be done slowly to ensure accurate readings. 7. When the switch is open, "OPEN" is shown on the display. Bleed the pump slowly until the pressure switch closes. Pressure Calibrator Pressure Switch Test At the top of the display "SW OPENED AT" and the pressure that the switch was opened at is shown as in the figure below. hix044.eps hix043.eps 27 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664
Pressure Calibrator
Pressure Switch Test
27
Note
In switch-test mode the display update rate is
increased to help capture pressure input
changes. Even with this enhanced sample rate,
pressurization of the device under test should be
done slowly to ensure accurate readings.
7.
When the switch is open, “OPEN” is shown on the
display. Bleed the pump slowly until the pressure
switch closes.
hix043.eps
At the top of the display “SW OPENED AT” and the
pressure that the switch was opened at is shown as in the
figure below.
hix044.eps
Shop for Fluke products online at:
1.888.610.7664
www.
MyFlukeStore
.com