Fluke 719PRO-300G Product Manual - Page 67
Pump_limit?, Remote, Rtd_type, Rtd_type?, St_start
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 67 highlights
PUMP_LIMIT? Returns the pump limit. The above example with PUMP_LIMIT? returns: 50.000 REMOTE Puts the Product into remote mode. From the remote mode, the keypad can be used to get back to local mode unless the command LOCKOUT was entered before REMOTE. Then the keypad is locked out, and the LOCAL command must be sent to get back to local operation. RTD_TYPE Sets the RTD type. The subsequent list shows RTD types the way they should be entered after the command: PT385_100; PT392_100; PTJIS_100; CUSTOM; For Example: RTD_TYPE PT385_100 sets RTD type to PT100-385 Pressure Calibrator Remote Operation RTD_TYPE? Returns the RTD type. For example: If the RTD type is PT385_100, RTD_TYPE?, PT100_385 is returned. SIM Sets the output for current simulation. This command also switches the Product into mA simulation mode. A number and a unit must be entered after the command. For example: SIM 5 MA sets the current simulation at 5 mA SIM? Returns the output of the current simulation. With the example above, the output would be: 5.000000E-03, A ST_START Starts a switch test. 57 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664