Custody Transfer / Flow Calibration, Remote Operation, Remote Interface
View all Fluke 719PRO-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 49 highlights
Custody Transfer / Flow Calibration The Product is ideal for flow computer calibration. Every manufacturer of flow computers has a different calibration procedure, but most call for calibration of three parameters: static pressure, differential pressure and temperature. To facilitate these measurements, recall setup #5 on the Product. 1. Connect the Product to the static and differential pressures. ([P1], EXT) Then connect the RTD sensor to the Product. 2. With the reading of the RTD, static, and differential pressures, make sure the flow computer has the correct reading. If not, adjust the flow computer as necessary. Pressure Calibrator Custody Transfer / Flow Calibration Remote Operation Remote Interface The Product can be remotely controlled with a PC terminal or by a computer program that runs the Product in an automated system. It uses an RS-232 serial port connection for remote operation. Note To use the remote control option, a custom miniature circular to USB interface cable, PN 4401616 must be purchased separately. To contact Fluke refer to the "Contact Fluke" section. With this connection the user can write programs on the PC, with Windows languages like Visual Basic to operate the Product, or use a terminal program similar to Hyper Terminal, to enter single commands. Typical RS-232 remote configurations are shown in Figure 11. 39 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664