Fluke 1586A/2HC Product Manual - Page 120

Fluke see

Page 120 highlights

1586A Users Manual Error Code Error 109 Error 110 Error 111 Error 200 Error 201 Error 202 Error 203 Error 204 Table 7-1. Error Messages (cont.) Error Message Unknown ACK byte Quality indicator too low GX restart A/D conversion timeout Error detecting line frequency A2 memory failed Slot 1 module memory failed Slot 2 module memory failed Causes and Solutions Cause: A packet was received, but it is not an acknowledgement. Solution: No user action necessary. The Instrument will restart the GX. Cause: GX quality low. Solution: No user action necessary. The Instrument will restart the GX. Cause: GX not responding. Solution: No user action necessary. Instrument will restart the GX. Cause: A/D conversion timeout. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: The line frequency is out of range. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on A2 board, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on module of slot 1, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on module of slot 2, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). 7-6 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

1586A
Users Manual
7-6
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 109
Unknown ACK byte
Cause:
A packet was received, but it is not an
acknowledgement.
Solution:
No user action necessary. The Instrument will
restart the GX.
Error 110
Quality indicator too
low
Cause:
GX quality low.
Solution:
No user action necessary. The Instrument will
restart the GX.
Error 111
GX restart
Cause:
GX not responding.
Solution:
No user action necessary. Instrument will restart
the GX.
Error 200
A/D conversion
timeout
Cause:
A/D conversion timeout.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 201
Error detecting line
frequency
Cause:
The line frequency is out of range.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 202
A2 memory failed
Cause:
Failed to access memory on A2 board, or checksum
failure detected in the memory.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 203
Slot 1 module
memory failed
Cause:
Failed to access memory on module of slot 1, or
checksum failure detected in the memory.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 204
Slot 2 module
memory failed
Cause:
Failed to access memory on module of slot 2, or
checksum failure detected in the memory.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).