Fluke 1586A/2HC Product Manual - Page 120
Fluke see
View all Fluke 1586A/2HC manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 120 highlights
1586A Users Manual Error Code Error 109 Error 110 Error 111 Error 200 Error 201 Error 202 Error 203 Error 204 Table 7-1. Error Messages (cont.) Error Message Unknown ACK byte Quality indicator too low GX restart A/D conversion timeout Error detecting line frequency A2 memory failed Slot 1 module memory failed Slot 2 module memory failed Causes and Solutions Cause: A packet was received, but it is not an acknowledgement. Solution: No user action necessary. The Instrument will restart the GX. Cause: GX quality low. Solution: No user action necessary. The Instrument will restart the GX. Cause: GX not responding. Solution: No user action necessary. Instrument will restart the GX. Cause: A/D conversion timeout. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: The line frequency is out of range. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on A2 board, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on module of slot 1, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Failed to access memory on module of slot 2, or checksum failure detected in the memory. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). 7-6 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664