Fluke 1586A/2HC Product Manual - Page 123

External CJC

Page 123 highlights

7 Error Messages and Troubleshooting Error Messages Error Code Error 399 Error 400 Error 401 Error 402 Error 403 Error 404 Error 405 Error 406 Table 7-1. Error Messages (cont.) Error Message Causes and Solutions Error at step %s Cause: An error occurred at present calibrating step. Solution: Check next error message for specific solution. External CJC not configured Cause: A thermocouple channel is set to external CJC, but the channel that is used to measure external CJC temperature has not been configured. Solution: Configure the lowest module channel to PRT or Thermistor with temperature reading, it will be used as external CJC channel. Channel used as external CJC Cause: Configure is changed to the channel that is used to measure external CJC temperature. Solution: Turn off all thermocouple channels that are configured for external CJC first. TC external reference channel Cause: Configure the lowest module channel as TC with external CJC. Solution: Check the parameter and make sure that the lowest module channel is not configured as TC with an external CJC. Parameter does not meet present config Cause: Parameter or command is not allowed for present configuration. For example: SENSe:NPLC 1 (@401) is received, but channel 401 is DIO. Solution: Check the parameter value and try again. Channel limit exceeded Cause: Too many channels in scan list (max. 178). Solution: Turn off some channels and try again. Channel used as alarm trigger Cause: A channel cannot be set to OFF while it is used as trigger channel for alarm test. Solution: Change the trigger channel setting or reconfigure the trigger source. Alarm channel not enabled Cause: The alarm channel selected to be the alarm trigger is set to OFF. Solution: Enable and configure the alarm channel then try again. 7-9 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

Error Messages and Troubleshooting
Error Messages
7
7-9
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 399
Error at step %s
Cause:
An error occurred at present calibrating step.
Solution:
Check next error message for specific solution.
Error 400
External CJC not
configured
Cause:
A thermocouple channel is set to external CJC, but
the channel that is used to measure external CJC
temperature has not been configured.
Solution:
Configure the lowest module channel to PRT or
Thermistor with temperature reading, it will be used as
external CJC channel.
Error 401
Channel used as
external CJC
Cause:
Configure is changed to the channel that is used to
measure external CJC temperature.
Solution:
Turn off all thermocouple channels that are
configured for external CJC first.
Error 402
TC external reference
channel
Cause:
Configure the lowest module channel as TC with
external CJC.
Solution:
Check the parameter and make sure that the
lowest module channel is not configured as TC with an
external CJC.
Error 403
Parameter does not
meet present config
Cause:
Parameter or command is not allowed for present
configuration. For example: SENSe:NPLC 1 (@401) is
received, but channel 401 is DIO.
Solution:
Check the parameter value and try again.
Error 404
Channel limit
exceeded
Cause:
Too many channels in scan list (max. 178).
Solution:
Turn off some channels and try again.
Error 405
Channel used as
alarm trigger
Cause:
A channel cannot be set to OFF while it is used as
trigger channel for alarm test.
Solution:
Change the trigger channel setting or reconfigure
the trigger source.
Error 406
Alarm channel not
enabled
Cause:
The alarm channel selected to be the alarm trigger is
set to OFF.
Solution:
Enable and configure the alarm channel then try
again.