3 Input and Channel Configuration Channel Configuration Thermocouple Channels Refer to Table 3-6 to configure a thermocouple channel. Thermocouple types R and S have optional correction polynomial coefficients to be entered. The correction equation is as follows: ΔV = C0 + C1 t + C2 t 2 + C3 t 3 where t is the temperature in °C and ΔV is the correction in millivolts. If the correction is not used, leave the coefficients set to 0. Table 3-6. Thermocouple Channel Configuration 1 2 6 3 4 5 Item Function Thermocouple function selection. Select the thermocouple type. Input a custom alpha-numeric label to help identify the channel (optional). Set an alarm for this channel (optional). See "HI and LO Channel Alarms" on page 3-31. Apply Mx+B scaling to the measurement (optional). See "Mx+B Scaling" on page 3-30. Load a thermocouple from the Probe Library. See "Probe Library" on page 3-34. hcn009.eps 3-21 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664