Fluke 1586A/2HC Product Manual - Page 62
Current and Voltage Channels
View all Fluke 1586A/2HC manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 62 highlights
3 Input and Channel Configuration Channel Configuration 4. After the channel is configured, verify that the channel is configured properly and reading the input. For instructions, see "Verify a Channel" on page 3-14. Current and Voltage Channels Refer to Table 3-4 to configure a dc voltage or current channel. Note Ch001 can be set up for all measurements. ChX21 and ChX22 are the only channels that can be set up for current measurements. To expand the functionality of the Product, current sources can also be connected to ChX01 through ChX20 by the use of an accessory shunt resistor. See Chapter 2 for part number and ordering information. Table 3-4. Current and Voltage Channel Configuration 1 2 3 4 5 Item Function DC voltage or DC current function selection. For voltage, set the voltage range to: Auto, 100 mV, 1 V, 10 V, or 50 V. For current, set the current range to: Auto, 100 μA, 1 mA, 10mA, or 100 mA. Input a custom alpha-numeric label to help identify the channel (optional). Set an alarm for this channel (optional). See "HI and LO Channel Alarms" on page 3-31. Apply Mx+B scaling to the measurement (optional). See "Mx+B Scaling" on page 3-30. hcn050.eps 3-19 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664