Fluke 1586A/2HC Product Manual - Page 133

Command DATA:LAST? or DATA:READ? was

Page 133 highlights

7 Error Messages and Troubleshooting Error Messages Error Code Error 1331 Error 1333 Error 1334 Error 1337 Error 1338 Error 1339 Error 1340 Error 1341 Error 1601 Error 1602 Error 1603 Table 7-1. Error Messages (cont.) Error Message Causes and Solutions Command incomplete Cause: An incomplete command line was received. Solution: Make sure that correct command was sent. Invalid channel number Cause: Invalid channel number was found by command parser. This could be caused by a wrong format. Solution: Check the parameter and try again. Illegal channel count Cause: The number of channels exceeds the limit when parse the channel list. Solution: Check the parameter and try again. Operation not allowed Cause: Operation is not allowed for current configuration, and the channel is not in scan list. Make sure that you have input the command with correct parameters and channel ID. Solution: Check the parameter and try again. Operation not allowed in DMM mode Cause: Operation is not allowed in DMM mode. Solution: Use command *RST to reset to factory default. Memory is empty Cause: Command DATA:LAST? or DATA:READ? was received, but data is not available. Solution: Check the trigger settings, and make sure that the Instrument is properly triggered for scan. File name too long Cause: More than 30 characters were found in filename which specified in command MEM:LOG:READ? or MMEM:LOG:READ? Solution: Check the parameter and try again. Scan sweep in process Cause: Operation not allowed while scan is in progress. Solution: Send command ABORt to stop the scan. Error copying file Cause: File coping failed while running SYSTem:DFU command because source is missing. Solution: Check to see if the file is in the memory. Error modifying file property Cause: Failed to change file property while running SYSTem:DFU command. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). USB memory not ready Cause: USB drive not ready for recording or memory operations. Solution: Install the USB drive. See Chapter 2. 7-19 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

Error Messages and Troubleshooting
Error Messages
7
7-19
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 1331
Command
incomplete
Cause:
An incomplete command line was received.
Solution:
Make sure that correct command was sent.
Error 1333
Invalid channel
number
Cause:
Invalid channel number was found by command
parser. This could be caused by a wrong format.
Solution:
Check the parameter and try again.
Error 1334
Illegal channel count
Cause:
The number of channels exceeds the limit when
parse the channel list.
Solution:
Check the parameter and try again.
Error 1337
Operation not
allowed
Cause:
Operation is not allowed for current configuration, and
the channel is not in scan list. Make sure that you have input
the command with correct parameters and channel ID.
Solution:
Check the parameter and try again.
Error 1338
Operation not
allowed in DMM
mode
Cause:
Operation is not allowed in DMM mode.
Solution:
Use command *RST to reset to factory default.
Error 1339
Memory is empty
Cause:
Command DATA:LAST? or DATA:READ? was
received, but data is not available.
Solution:
Check the trigger settings, and make sure that the
Instrument is properly triggered for scan.
Error 1340
File name too long
Cause:
More than 30 characters were found in filename
which specified in command MEM:LOG:READ? or
MMEM:LOG:READ?
Solution:
Check the parameter and try again.
Error 1341
Scan sweep in
process
Cause:
Operation not allowed while scan is in progress.
Solution:
Send command ABORt to stop the scan.
Error 1601
Error copying file
Cause:
File coping failed while running SYSTem:DFU
command because source is missing.
Solution:
Check to see if the file is in the memory.
Error 1602
Error modifying file
property
Cause:
Failed to change file property while running
SYSTem:DFU command.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 1603
USB memory not
ready
Cause:
USB drive not ready for recording or memory
operations.
Solution:
Install the USB drive. See Chapter 2.