Fluke 1586A/2HC Product Manual - Page 84
Con a Scan
View all Fluke 1586A/2HC manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 84 highlights
1586A Users Manual • The total channel sample time is the minimum sample time plus the userprogrammed channel delay. The longer the channel delay, the longer it takes to sample the channel. • The total scan sweep time is sum of the channel sample times of all the channels scanned. A scan sweep will start when the selected trigger source is asserted if a previous scan sweep is not still in progress (see "Trigger Type" on page 4-7). • The interval at which scan sweeps occur can be fixed by programming the trigger type to Timer and setting an interval time. SCAN START Channel List Channel 102 Channel 104 Channel 115 Channel Delay Set Channel 117 Channel Delay Set Channel 121 Key: - Sample Time* - Channel Delay *Consists of NPLC time, settling time, and ADC samples. Scan Cycle 1 Remaining Interval Scan Cycle 2 RReemmaaiinniinngg IInntteerrvvaall Sweep Time Scan Interval Sweep Time Scan Interval SCAN FINISH Figure 4-2. Illustration of a Scan Sweep hcn057.eps Configure a Scan Scans are configured in the Test Setup menu that is opened in the Channel Setup menu (push then push ). This section contains information on each parameter in the Test Setup menu to help configure the scan. Figure 4-3 shows the Test Setup menu. 4-6 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664