Seat-To-Floor Height Adjustment, Adjusting Forks
View all Invacare SOLARA3G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 87 highlights
12 12.3 Seat-To-Floor Height Adjustment ƽ WARNING Repositioning the axle plates (seat-to-floor height) MUST be performed by a qualified technician. WHEELS/FORKS 12.4 Adjusting Forks 1. Remove the dust cover from the caster headtube. 2. To properly tighten caster journal system and guard against flutter, perform the following check: A. Tip back the wheelchair to floor. B. Pivot both forks and casters to top of their arc simultaneously. C. Let casters drop to bottom of arc (wheels should swing once to one-side, then IMMEDIATELY rest in a straight downward position). D. Adjust locknuts according to freedom of caster swing. 3. Test wheelchair for maneuverability. 4. Readjust locknuts if necessary, and repeat STEPS 1-3 until correct. 5. Snap dust cover into the caster headtube. 6. Repeat STEPS 1-5 for opposite side, if necessary. Locknut Dust Cover Caster Headtube FIGURE 12.2 Adjusting Forks Part No 1154295 87 Invacare® Solara®3G/Spree 3G Wheelchair