Fluke 1625 FE 1625 Users Manual - Page 21
Measurement of Interference Voltage DC + AC (UST), Earthing Resistance (RE), Measuring, Range, Display - 2 earth ground tester
View all Fluke 1625 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 21 highlights
Earth/Ground Tester Specifications Measurement of Interference Voltage DC + AC (UST) Measuring method: fullwave rectification Measuring Range Display Range Resolution Frequency Range Limits of Error 1...50 V 0.0...50 V 0.1 V DC/AC 45...400 Hz sine ± (5% of reading +5 digit) Measuring sequence: Internal resistance: Max. overload: approx. 4 measurements /s approx. 1.5 MΩ Urms = 250 V Measurement of Interference Frequency (FST) Measuring method: Measurement of oscillation period of the interference voltage Measuring Range 16.0 ... 400 Hz Display Range Resolution Range 16.0...299.9...99 9 Hz 0.1 ... 1 Hz 1 V ... 50 V Limits of Error ± (1% v. mv +2 digit) Earthing Resistance (RE) Measuring method: current and voltage measurement with probe as IEC61557-5 Open circuit voltage: 20 / 48 V, AC Short circuit current: 250 mA AC Measuring frequency: 94 , 105 , 111 , 128 Hz selected manually or automatic.(AFC) 55 Hz in function R* Noise rejection: 120 dB ( 16 2/3, 50 , 60, 400 Hz ) Max. overload: Urms = 250 V 13