CPI Communications
![]() |
View all Campbell Scientific CSAT3B manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 46 highlights
CSAT3B Three-Dimensional Sonic Anemometer FIGURE 7-11. SDM star topology (CSAT3B sensors arms and grounding cables not shown) 7.7.2 CPI Communications If data collection from the anemometer is to be done by a datalogger using CPI communications, then connect a CSAT3BCBL2 to the Power/SDM port by screwing in the M16 connector into the port until tight. NOTE A CSAT3BCBL1 may also be used for power, but the SDM wires will not be used. The CSAT3BCSBL1 also has lighter gauge power wires, which increases voltage loss over long cable length, so it should only be used for shorter cable lengths. Next, connect a CSAT3BCBL3 to the CPI/RS-485 port in the same manner. If only one CSAT3B is being measured, the opposite end of the power cable should have pigtail wire leads which may be wired to the 12 V and G terminals on a datalogger or to another 12 to 32 Vdc power supply. The CPI cable should have the RJ-45 connector plugged into the CPI port of the datalogger. 36