Fluke 1586A/2DS Product Manual - Page 112
Memory Reset and Factory Reset
View all Fluke 1586A/2DS manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 112 highlights
1586A Users Manual 1 2 Figure 6-1. Fuse Replacement hcn027.eps Memory Reset and Factory Reset The Product has two memory reset functions to remove data from the memory and to reset the Product: Clear all Files and Factory Reset. See Table 6-2 for a comparison of the three functions. Note All memory reset functions require the Admin password to accomplish. Table 6-2. Comparison of the Memory Clear Functions Task Clear All Files Factory Reset Deletes Test Setup files, DMM Data files, and Scan Data files from the internal memory. [1] • Clears the configuration of the Channel Setup, Test Setup, and Instrument Setup [2] • Notes: [1] - Does not remove data from the USB drive. [2] - Does not reset the MAC address, the serial number, calibration, clock time, nor the Admin or User passwords. 6-4 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664