Fluke 1586A/2DS Product Manual - Page 94
Automated Test
View all Fluke 1586A/2DS manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 94 highlights
1586A Users Manual Automated Test Overview The Product has an automated test feature that works directly with an external temperature source (the source) to calibrate and verify the performance of temperature probes (UUT) against a calibrated reference probe. To fully automate the test, the user can connect to and control a temperature source such as dry-wells, microbaths, or fluid baths from Fluke Calibration. In this configuration, the Product communicates the setpoint to the source via the RS-232 connection on the rear panel (see Figure 4-7). The source is then activated and sets the temperature to the setpoint. When the measurement from the reference temperature probe that is wired to the Product meets all the parameters of the setpoint (tolerance, stability, and soak time), the Product automatically communicates the next setpoint. This sequence repeats until all setpoints are complete. Note The Automated Test has a Control Source setting that can be toggled to On or Off. When Off, the operator must manually set the temperature on the source. Temperature Source Control Output Reference Probe UUT Figure 4-7. External Temperature Control Source Example Temperature Control Source hcn042.eps 4-16 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664