7 Error Messages and Troubleshooting Error Messages Error Code Error 207 Error 208 Error 209 Error 210 Error 211 Error 260 Error 300 Error 301 Error 302 Error 303 Error 304 Error 305 Table 7-1. Error Messages (cont.) Error Message Causes and Solutions Failed to reset relay on slot 1 Failed to reset relay on slot 2 Failed to reset relay on slot 3 Cause: The relay does not reset when the channel is switched. Solution: Cycle the power and try again. If the error occurs again, the relay card needs serviced, contact Fluke (see Chapter 1). Error reading cold junction temperature (Slot 1) Cause: The CJC temperature cannot be read from external module of slot 1. Solution: Cycle the power. If the error occurs again, the relay card needs serviced, contact Fluke (see Chapter 1). Error reading cold junction temperature (Slot 2) Cause: The CJC temperature cannot be read from external module of slot 2. Solution: Cycle the power. If the error occurs again, the relay card needs serviced, contact Fluke (see Chapter 1). Error copying configure to ChXXX Cause: The copy function failed. Solution: Check the channel list and make sure all selected channel are set up correctly. Invalid procedure Cause: Wrong calibration procedure is specified in CALibrate:STARt command or MAIN in CALibrate:STARt command could be misspelled. Solution: Check the parameter value and try again. Invalid procedure step Cause: Wrong or misspelled calibration step is specified in CALibrate:STARt command. Solution: Check the parameter value and try again. Step change not allowed while busy Cause: Invalid CALibrate:STARt was received while a calibrate operation is in progress. Solution: Send CALibrate:ABORt to terminate the active calibration procedure. Invalid starting point Cause: The step specified in CALibrate:STARt command should be the first point of a function or a range, it is invalid to start a calibration procedure from an intermediate step. Solution: Check the start point and try again. Invalid reference unit Cause: CALibrate:NEXT command was received with an unexpected unit. Solution: Check the parameter value and try again. Entered value out of range Cause: CALibrate:NEXT command was received, but the reference value is out of range. Solution: Check the parameter value and try again. 7-7 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

Error Messages and Troubleshooting
Error Messages
7
7-7
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 207
Failed to reset relay
on slot 1
Cause:
The relay does not reset when the channel is
switched.
Solution:
Cycle the power and try again. If the error occurs
again, the relay card needs serviced, contact Fluke (see
Chapter 1).
Error 208
Failed to reset relay
on slot 2
Error 209
Failed to reset relay
on slot 3
Error 210
Error reading cold
junction temperature
(Slot 1)
Cause:
The CJC temperature cannot be read from external
module of slot 1.
Solution:
Cycle the power. If the error occurs again, the relay
card needs serviced, contact Fluke (see Chapter 1).
Error 211
Error reading cold
junction temperature
(Slot 2)
Cause:
The CJC temperature cannot be read from external
module of slot 2.
Solution:
Cycle the power. If the error occurs again, the relay
card needs serviced, contact Fluke (see Chapter 1).
Error 260
Error copying
configure to ChXXX
Cause:
The copy function failed.
Solution:
Check the channel list and make sure all selected
channel are set up correctly.
Error 300
Invalid procedure
Cause:
Wrong calibration procedure is specified in
CALibrate:STARt command or MAIN in CALibrate:STARt
command could be misspelled.
Solution:
Check the parameter value and try again.
Error 301
Invalid procedure
step
Cause:
Wrong or misspelled calibration step is specified in
CALibrate:STARt command.
Solution:
Check the parameter value and try again.
Error 302
Step change not
allowed while busy
Cause:
Invalid CALibrate:STARt was received while a
calibrate operation is in progress.
Solution:
Send CALibrate:ABORt to terminate the active
calibration procedure.
Error 303
Invalid starting point
Cause:
The step specified in CALibrate:STARt command
should be the first point of a function or a range, it is invalid to
start a calibration procedure from an intermediate step.
Solution:
Check the start point and try again.
Error 304
Invalid reference unit
Cause:
CALibrate:NEXT command was received with an
unexpected unit.
Solution:
Check the parameter value and try again.
Error 305
Entered value out of
range
Cause:
CALibrate:NEXT command was received, but the
reference value is out of range.
Solution:
Check the parameter value and try again.