Fluke 1586A/2DS Product Manual - Page 131

Error reading file

Page 131 highlights

7 Error Messages and Troubleshooting Error Messages Error Code Error 825 Error 826 Error 827 Error 828 Error 901 Error 903 Error 904 Error 905 Error 906 Error 907 Error 908 Error 909 Table 7-1. Error Messages (cont.) Error Message LAN address invalid LAN hostname invalid LAN hostname too long LAN connection timeout File not found Error exporting file Error deleting file Error reading file Error defining setup file name Error reading folder Error writing file File already exists Causes and Solutions Cause: Invalid IP, mask or gateway address was found. Solution: Check the address and try again. Cause: The hostname contains invalid character. The Instrument only allows 'A' to 'Z', '0' to '9' and '-'. Solution: Check the parameter and try again. Cause: More than 20 characters were found when configuring hostname. Solution: Check the parameter and try again. Cause: The LAN connection has been disconnected. Solution: Check the LAN equipment and try to connect again. Cause: File not found for memory operation. Solution: Check to make sure the file is in the memory. Cause: Failed while exporting file to USB memory. Solution: Make sure that the USB memory is not write protected and that the USB memory has enough free space. Cause: Failed while deleting file. Solution: If you are delete file from USB memory, make sure that the USB memory is not write-protected. Cause: The file has been damaged. Solution: Cycle the power. If the error occurs again, contact Fluke (see Chapter 1). Cause: Duplicate filename was found. Solution: Try another name or remove the file with the conflict. Cause: Folder not exist. Solution: Make sure that the folder is in the memory. Cause: Failed while writing file. Solution: Make sure that the USB memory is not write protected and that the USB memory has enough free space. Cause: The destination already exists while a file with the same name. Solution: Try another name or remove the file with the conflict. 7-17 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

Error Messages and Troubleshooting
Error Messages
7
7-17
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 825
LAN address invalid
Cause:
Invalid IP, mask or gateway address was found.
Solution:
Check the address and try again.
Error 826
LAN hostname
invalid
Cause:
The hostname contains invalid character. The
Instrument only allows ‘A’ to ‘Z’, ‘0’ to ‘9’ and ‘-‘.
Solution:
Check the parameter and try again.
Error 827
LAN hostname too
long
Cause:
More than 20 characters were found when
configuring hostname.
Solution:
Check the parameter and try again.
Error 828
LAN connection
timeout
Cause:
The LAN connection has been disconnected.
Solution:
Check the LAN equipment and try to connect
again.
Error 901
File not found
Cause:
File not found for memory operation.
Solution:
Check to make sure the file is in the memory.
Error 903
Error exporting file
Cause:
Failed while exporting file to USB memory.
Solution:
Make sure that the USB memory is not write
protected and that the USB memory has enough free space.
Error 904
Error deleting file
Cause:
Failed while deleting file.
Solution:
If you are delete file from USB memory, make sure
that the USB memory is not write-protected.
Error 905
Error reading file
Cause:
The file has been damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 906
Error defining setup
file name
Cause:
Duplicate filename was found.
Solution:
Try another name or remove the file with the
conflict.
Error 907
Error reading folder
Cause:
Folder not exist.
Solution:
Make sure that the folder is in the memory.
Error 908
Error writing file
Cause:
Failed while writing file.
Solution:
Make sure that the USB memory is not write
protected and that the USB memory has enough free space.
Error 909
File already exists
Cause:
The destination already exists while a file with the
same name.
Solution:
Try another name or remove the file with the
conflict.