Fluke 1586A/2DS Product Manual - Page 119

GX CRC error

Page 119 highlights

7 Error Messages and Troubleshooting Error Messages Error Code Error -102 Error 0 Error 100 Error 101 Error 102 Error 103 Error 104 Error 105 Error 106 Error 107 Error 108 Table 7-1. Error Messages (cont.) Error Message Causes and Solutions Syntax error No error Cause: Invalid syntax was found in the command string. Solution: Check the parameter value and try again. Cause: Error queue is empty. No action necessary. Acknowledgement queue full Cause: GX acknowledgement queue overflow. Note GX is the abbreviation for Guard Crossing - which is the communication link between the inguard and outguard. Solution: Instrument will restart the GX. Inguard not responding (recv) Cause: Inguard not responding (receive). Solution: Instrument will restart the GX. Lost sync with inguard Cause: Lost sync with inguard. Solution: Instrument will restart the GX. Inguard not responding (send) Cause: Inguard not responding (send). Solution: Instrument will restart the GX. Oversize packet received Cause: Oversize packet received. Solution: The packet will be ignored automatically. GX CRC error Cause: CRC error detected. Solution: No user action necessary. The packet will be ignored automatically. Wrong ACK number Cause: An ACK packet was received with unexpected ACK number. Solution: No user action necessary. The Instrument will restart the GX. Info packet received; link not active Cause: An INFO packet was received, but the communication has not been setup. Solution: No user action necessary. The packet will be ignored automatically. Cause: A packet was received with unknown control byte. Unknown control byte Solution: No user action necessary. Instrument will restart the GX. 7-5 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

Error Messages and Troubleshooting
Error Messages
7
7-5
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error -102
Syntax error
Cause:
Invalid syntax was found in the command string.
Solution:
Check the parameter value and try again.
Error 0
No error
Cause:
Error queue is empty. No action necessary.
Error 100
Acknowledgement
queue full
Cause:
GX acknowledgement queue overflow.
Note
GX is the abbreviation for Guard Crossing - which is the
communication link between the inguard and outguard.
Solution:
Instrument will restart the GX.
Error 101
Inguard not
responding (recv)
Cause:
Inguard not responding (receive).
Solution:
Instrument will restart the GX.
Error 102
Lost sync with
inguard
Cause:
Lost sync with inguard.
Solution:
Instrument will restart the GX.
Error 103
Inguard not
responding (send)
Cause:
Inguard not responding (send).
Solution:
Instrument will restart the GX.
Error 104
Oversize packet
received
Cause:
Oversize packet received.
Solution:
The packet will be ignored automatically.
Error 105
GX CRC error
Cause:
CRC error detected.
Solution:
No user action necessary. The packet will be
ignored automatically.
Error 106
Wrong ACK number
Cause:
An ACK packet was received with unexpected ACK
number.
Solution:
No user action necessary. The Instrument will
restart the GX.
Error 107
Info packet received;
link not active
Cause:
An INFO packet was received, but the
communication has not been setup.
Solution:
No user action necessary. The packet will be
ignored automatically.
Error 108
Unknown control byte
Cause:
A packet was received with unknown control byte.
Solution:
No user action necessary. Instrument will restart
the GX.