Fluke 1586A/2DS Product Manual - Page 58
Cancel
View all Fluke 1586A/2DS manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 58 highlights
3 Input and Channel Configuration Channel Configuration 2. Push . 3. Push or to highlight a channel. 4. Push to edit the channel. 5. Push to open the Verify Channel menu. 6. Push to zero the channel. A confirmation message appears, select OK to continue or Cancel to disregard the changes and return to previous menu. After OK is selected, the Product sets the Mx+B to ON and loads the calculated offset into the Mx+B offset setting. To clear the zero function, set the Mx+B back to the original values. See "Mx+B Scaling" on page 3-30 for instructions. Mx+B Indicator Relative Measurement Value Baseline Measurement Zero Function Softkey Figure 3-7. Zero Function hcn060.eps Copy a Channel The Product has a copy and paste function to help duplicate channels. When a channel is duplicated, all of the channel configuration settings are copied and pasted to a new channel number. The copy channel softkey in located in the Channel Setup menu. Note When a channel is copied to another channel, the channel configuration of the target channel is overwritten without a notification. Confirm the channel selections before OK is pushed to prevent accidental overwrites. To duplicate a channel with copy and paste: 1. Push . 2. Push or to highlight a channel. 3-15 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664