Fluke 1586A/2DS Product Manual - Page 88

Temperature Unit, Align Channels

Page 88 highlights

1586A Users Manual Temperature Unit This is an overall Product setting that sets the temperature units to be shown in either Celsius or Fahrenheit. Notes • The ability to change this setting to Fahrenheit is not available in some regions. • Mx+B and Alarm settings of temperature channels are reset when the temperature unit is changed. Align Channels The Align Channels function lets the user designate a reference channel to align other channels to so that measured offsets between sensors can be set to zero relative to one another and the reference channel. This is accomplished with an Mx+B offset that is automatically applied to the selected channels to make them match the reference channel. An application where this is commonly used is for temperature measurements. For example, a reference probe that is fully calibrated is put into a chamber with other probes that reads 250.52 °C. With the align function, the user sets the calibrated reference probe channel to be the reference channel. The user then sets four probes channels to be aligned with that reference that currently read 250.52 °C, 250.68 °C, 250.71 °C, 250.33 °C. When the user aligns the probes, the Product calculates and applies an Mx+B offset to the probes so that the displayed measurement value is the same as the reference probe. The result is that all four probes now measure 250.52 °C. To align to a reference channel: 1. Push . 2. Push . 3. Push . 4. Use  and  to select a channel as the reference then push . 5. Use  and  to highlight a channel then push  to select or deselect. Multiple channels can be selected to align. 6. Push  to align the channels. The Product will take measurements on the reference channel and selected channels, then use Mx+B function to align these channels to the reference. 4-10 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135

1586A
Users Manual
4-10
Temperature Unit
This is an overall Product setting that sets the temperature units to be shown in either
Celsius or Fahrenheit.
Notes
The ability to change this setting to Fahrenheit is not available in
some regions.
Mx+B and Alarm settings of temperature channels are reset when the
temperature unit is changed.
Align Channels
The Align Channels function lets the user designate a reference channel to align other
channels to so that measured offsets between sensors can be set to zero relative to one
another and the reference channel. This is accomplished with an Mx+B offset that is
automatically applied to the selected channels to make them match the reference channel.
An application where this is commonly used is for temperature measurements. For
example, a reference probe that is fully calibrated is put into a chamber with other probes
that reads 250.52 °C. With the align function, the user sets the calibrated reference probe
channel to be the reference channel. The user then sets four probes channels to be aligned
with that reference that currently read 250.52 °C, 250.68 °C, 250.71 °C, 250.33 °C. When
the user aligns the probes, the Product calculates and applies an Mx+B offset to the
probes so that the displayed measurement value is the same as the reference probe. The
result is that all four probes now measure 250.52 °C.
To align to a reference channel:
1.
Push
.
2.
Push
.
3.
Push
.
4.
Use
and
to select a channel as the reference then push
.
5.
Use
and
to highlight a channel then push
to select or deselect. Multiple
channels can be selected to align.
6.
Push
to align the channels. The Product will take measurements on the
reference channel and selected channels, then use Mx+B function to align these
channels to the reference.