Fluke 726 FE 726 Users Manual - Page 25
Shut Down Mode, CJC, Celcius and Fahrenheit (°C and °F), Frequency Pulse Output Voltage
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
Shut Down Mode The Calibrator comes with a shut down mode set for 30 minutes (displayed for about 1 second when the Calibrator is initially turned on). When shut down mode is enabled, the Calibrator automatically shuts down after the elapsed time from when the last key was pressed. 1. Press C until SHUT DOWN appears on the display. 2. Use X and W to increase or decrease the time. 3. Press S to save the setting. CJC Cold Junction Compensation (CJC) is a value for the cold end of a thermocouple at the Meter's end. 1. Press C until SELECT CJC appears on the display. 2. Use Y and Z to select ON or OFF. 3. Press S to save the setting. Celcius and Fahrenheit (°C and °F) 1. Press C until SELECT UNIT °C (or °F) appears on the display. 2. Use X and W to select °C or °F. 3. Press S to save the setting. Multifunction Process Calibrator Configuration Menus Frequency Pulse Output Voltage 1. Press C until FREQ OUTPUT V Adjust appears on the display. 2. Use X, W, Y and Z to adjust the frequency pulse output voltage from 1 to 20 V. 3. Press S to save the setting. Pulse Output Frequency 1. Press C until PULSE OUTPUT Hz FREQ Adjust appears on the display. 2. Use X, W, Y and Z to adjust the pulse output frequency from 2 CPM to 15 kHz. 3. Press S to save the setting. 15