Using Source Mode, Sourcing 4 to 20 mA, Simulating a 4- to 20-mA Transmitter
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 40 highlights
726 Users Manual Using Source Mode In SOURCE mode, the Calibrator: • generates calibrated signals for testing and calibrating process instruments • supplies voltages, currents, frequencies, and resistances • simulates the electrical output of RTD and thermocouple temperature sensors • measures gas pressure from an external source, creating a calibrated pressure source. Sourcing 4 to 20 mA To select the current sourcing mode, proceed as follows: 1. Connect the test leads in the mA terminals (left column). M 2. If necessary, press for SOURCE mode. 3. Press V for current and enter the desired current by pressing X,W,Y, and Z. Simulating a 4- to 20-mA Transmitter Simulate is a special mode of operation in which the Calibrator is connected into a loop in place of a transmitter and supplies a known, settable test current. Proceed as follows: 1. Connect the 24 V loop power source as shown in Figure 14. M 2. If necessary, press for SOURCE mode. 3. Press V until both mA and SIM display. 4. Enter the desired current by pressing X,W,Y, and Z. 30