Calibrating an I/P Device The following steps explain how to calibrate a device that controls pressure. Proceed as follows: 1. Connect the test leads to the instrument under test as shown in Figure 22. The connections simulate a current-to-pressure transmitter and measures the corresponding output pressure. 2. Press A (upper display). Multifunction Process Calibrator Calibrating an I/P Device 3. Press V for sourcing current (lower display). M 4. If necessary, press for SOURCE mode. 5. Enter the desired current by pressing X and W. Press Y and Z to select different digits. 49
Multifunction Process Calibrator
Calibrating an I/P Device
49
Calibrating an I/P Device
The following steps explain how to calibrate a device that
controls pressure. Proceed as follows:
1.
Connect the test leads to the instrument under test as
shown in Figure 22. The connections simulate a
current-to-pressure transmitter and measures the
corresponding output pressure.
2.
Press
A
(upper display).
3.
Press
V
for sourcing current (lower display).
4.
If necessary, press
M
for SOURCE mode.
5.
Enter the desired current by pressing
X
and
W
.
Press
Y
and
Z
to select different digits.