Fluke 726 FE 726 Users Manual - Page 54
Recall Data, Pulse Train Source/Read
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 54 highlights
726 Users Manual Memory Locations Save Data Mode Indicator S A V E D A T A Data Point 1 bec46f.eps Figure 19. SAVE DATA Menu Showing Measurement Memory Location 3, 1 Recall Data To recall data: 1. Push S twice. The right side of the display changes to show RECL SETUP and RECALL DATA. 2. Push W to highlight RECL DATA (bottom right of the display). 3. Push S. 4. Push Z to choose the desired memory location (top of the display). The data saved in that first memory location now appears. There can be different measurements stored (1-5) for each memory location (1-8). 44 5. Push Y or Z to select the correct data location (bottom right of display). 6. Push S to recall the data stored in that location. 7. Push W to return to the same RECALL DATA location to see the next saved measurement, 2 of 5 for example. Pulse Train Source/Read Pulse Train Source/Read counts input pulses or sources output pulses. Use the configuration menus to set the frequency and output voltage. Refer to "Configuration Menus" earlier in this manual. The number of counts is set through the main display and cannot be changed while sourcing pulses. L works as a trigger/stop key in this mode since ramping or stepping during a pulse train is not relevant.