RTD Accuracy (Read and Source) (ITS-90), Range, Minimum, Accuracy
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 73 highlights
Multifunction Process Calibrator Specifications RTD Accuracy (Read and Source) (ITS-90) Range Minimum Maximum Accuracy Ni120 (672) -80.00 260.00 0.15 -200.00 100.00 0.15 100.00 300.00 0.25 Pt100 (385) 300.00 600.00 0.35 600.00 800.00 0.45 -200.00 100.00 0.15 Pt100 (3926) 100.00 300.00 0.25 300.00 630.00 0.35 -200.00 100.00 0.15 Pt100 (3916) 100.00 300.00 0.25 300.00 630.00 0.35 -200.00 100.00 0.75 Pt200 (385) 100.00 300.00 0.85 300.00 630.00 0.95 -200.00 100.00 0.35 Pt500 (385) 100.00 300.00 0.45 300.00 630.00 0.55 -200.00 100.00 0.15 Pt1000 (385) 100.00 300.00 0.25 300.00 630.00 0.35 CU10 -10.00 250.00 1.8 Notes: Read Accuracy is based on 4-wire input. For 3-wire input, add ± 0.05 e assuming all three RTD leads are matched. Source Accuracy is based on 0.5 to 3.0 mA excitation current (0.1 mA for pt1000 range) 63