Fluke 726 FE 726 Users Manual - Page 69
Specifications, DC Voltage Measurement and Source
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 69 highlights
Multifunction Process Calibrator Specifications Specifications Specifications are based on a one year calibration cycle and apply from +18 °C to +28 °C unless stated otherwise. All specifications assume a five-minute warmup period. DC Voltage Measurement and Source Range Minimum Maximum Accuracy, (% of Reading + Floor) 30 V (upper display) 20 V (lower display) 20 V (Source) 100 mV (Source) 90 mV (Read) 0.000 0.000 0.000 0.000 0.000 30.000 20.000 20.000 100.000 90.000 0.010 % + 2 mV 0.010 % + 2 mV 0.010 % + 2 mV 0.010 % + 10 µV 0.010 % + 10 µV Maximum current output in voltage ranges is 1 mA with an output impedance of ≤ 1 Ω DC mA Measurement and Source Range Minimum mA Read (Upper Display) mA Read (Lower Display) mA Source 0.000 0.000 0.000 Maximum 24.000 24.000 24.000 Accuracy, (% of Reading + Floor) 0.010 % + 2 µA 0.010 % + 2 µA 0.010 % + 2 µA Maximum load on, mA source is 1 ke. With the HART resistor on, maximum load is 750 Ω. Voltage input range on simulate mode is 5 to 30 V 59