Fluke 726 FE 726 Users Manual - Page 28

Using Measure Mode, Measuring Electrical Parameters (Upper Display)

Page 28 highlights

726 Users Manual Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, or to measure the output of a 700 Series pressure module, use the upper display and proceed as follows: 1. Press l to select volts or current. LOOP should not be on. 2. Connect the leads as shown in Figure 7. Current Measurement with Loop Power The loop power function activates a 24 V supply in series with the current measuring circuit, allowing you to test a transmitter when it is disconnected from plant wiring. To measure current with loop power: 1. Connect the Calibrator to the transmitter current loop terminals as shown in Figure 8. 2. Press l while the Calibrator is in current measurement mode. LOOP appears and an internal 24 V loop supply turns on. Note When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels. 18 bec42f.eps Figure 7. Measuring Voltage and Current Output

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

726
Users Manual
18
Using Measure Mode
Measuring Electrical Parameters (Upper Display)
To measure the current or voltage output of a transmitter,
or to measure the output of a 700 Series pressure
module, use the upper display and proceed as follows:
1.
Press
l
to select volts or current. LOOP should
not be on.
2.
Connect the leads as shown in Figure 7.
Current Measurement with Loop Power
The loop power function activates a 24 V supply in series
with the current measuring circuit, allowing you to test a
transmitter when it is disconnected from plant wiring. To
measure current with loop power:
1.
Connect the Calibrator to the transmitter current loop
terminals as shown in Figure 8.
2.
Press
l
while the Calibrator is in current
measurement mode. LOOP appears and an internal
24 V loop supply turns on.
Note
When HART resistor mode is selected, the
250
resistor is turned on both mA channels.
bec42f.eps
Figure 7. Measuring Voltage and Current Output