Fluke 726 FE 726 Users Manual - Page 28
Using Measure Mode, Measuring Electrical Parameters (Upper Display)
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 28 highlights
726 Users Manual Using Measure Mode Measuring Electrical Parameters (Upper Display) To measure the current or voltage output of a transmitter, or to measure the output of a 700 Series pressure module, use the upper display and proceed as follows: 1. Press l to select volts or current. LOOP should not be on. 2. Connect the leads as shown in Figure 7. Current Measurement with Loop Power The loop power function activates a 24 V supply in series with the current measuring circuit, allowing you to test a transmitter when it is disconnected from plant wiring. To measure current with loop power: 1. Connect the Calibrator to the transmitter current loop terminals as shown in Figure 8. 2. Press l while the Calibrator is in current measurement mode. LOOP appears and an internal 24 V loop supply turns on. Note When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels. 18 bec42f.eps Figure 7. Measuring Voltage and Current Output