Fluke 726 FE 726 Users Manual - Page 52
Manually Stepping the mA Output, Auto Ramping the Output, Storing and Recalling Setups, Store a Setup
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 52 highlights
726 Users Manual Manually Stepping the mA Output To manually step current output: • Use H or I to step the current up or down in 25 % steps. • Touch either J to go to 0 %, or G to go to 100 %. Auto Ramping the Output Auto ramping can continuously apply a varying stimulus from the Calibrator to a transmitter, while your hands remain free to test the response of the transmitter. When L is pressed, the Calibrator produces a repeating 0 % - 100 % - 0 % ramp in a choice of three ramp waveforms: • E 0 % - 100 % - 0 % 40-second smooth ramp • P 0 % - 100 % - 0 % 15-second smooth ramp • N 0 % - 100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step. Steps are listed in Table 7. 42 To exit ramping, press any button. Table 7. mA Step Values Step 4 to 20 mA 0 % 4.000 25 % 8.000 50 % 12.000 75 % 16.000 100 % 20.000 Storing and Recalling Setups Up to eight settings can be stored in a nonvolatile memory to recall for later use. A low battery condition or a battery change does not jeopardize the stored settings. Store a Setup To store a setup: 1. Create the desired setup. 2. Push S. The right side of the display changes to show SAVE SETUP and SAVE DATA. 3. Push Sto select SAVE SETUP. 4. Push Y or Z to select the desired memory location (at the top of the LCD). 5. Push S to enter the setup.