Fluke 726 FE 726 Users Manual - Page 55

Calibrating a Transmitter, M

Page 55 highlights

Calibrating a Transmitter Use the measurement (upper display) and source (lower display) modes to calibrate a transmitter. This section applies to all but pressure transmitters. The following example shows how to calibrate a temperature transmitter. Use the following steps to calibrate a transmitter: 1. Connect the Calibrator to the instrument under test as shown in Figure 20. 2. Press l for current (upper display). If required, press l again to activate loop power. 3. Press T (lower display). If desired, continue pressing this key to select the desired thermocouple type. Multifunction Process Calibrator Calibrating a Transmitter M 4. If necessary, press for SOURCE mode. 5. Set the zero and span parameters by pressing X and W. Enter these parameters by pressing and holding J and G. For more information on setting parameters, see "Setting 0 % and 100 % Output Parameters". 6. Perform test checks at 0-25-50-75-100 % points by pressing H or I. Adjust the transmitter as necessary. Note When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels. 45

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Multifunction Process Calibrator
Calibrating a Transmitter
45
Calibrating a Transmitter
Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. This section
applies to all but pressure transmitters. The following
example shows how to calibrate a temperature
transmitter. Use the following steps to calibrate a
transmitter:
1.
Connect the Calibrator to the instrument under test
as shown in Figure 20.
2.
Press
l
for current (upper display). If required,
press
l
again to activate loop power.
3.
Press
T
(lower display). If desired, continue
pressing this key to select the desired thermocouple
type.
4.
If necessary, press
M
for SOURCE mode.
5.
Set the zero and span parameters by pressing
X
and
W
. Enter these parameters by pressing and
holding
J
and
G
. For more information on
setting parameters, see “Setting 0 % and 100 %
Output Parameters”.
6.
Perform test checks at 0-25-50-75-100 % points by
pressing
H
or
I
. Adjust the transmitter as
necessary.
Note
When HART resistor mode is selected, the
250
resistor is turned on both mA channels.