Fluke 726 FE 726 Users Manual - Page 57

Calibrating a Pressure Transmitter, UM

Page 57 highlights

Calibrating a Pressure Transmitter The following steps explain calibrating a pressure transmitter. 1. Connect the Calibrator to the instrument under test as shown in Figure 21. 2. Press l for current (upper display). If required, press l again to activate loop power. 3. Press U (lower display). Multifunction Process Calibrator Calibrating a Pressure Transmitter M 4. If necessary, press for SOURCE mode. 5. Zero the pressure module. 6. Perform checks at 0 % and 100 % of span and adjust the transmitter as necessary. Note When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels. 47

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80

Multifunction Process Calibrator
Calibrating a Pressure Transmitter
47
Calibrating a Pressure Transmitter
The following steps explain calibrating a pressure
transmitter.
1.
Connect the Calibrator to the instrument under test
as shown in Figure 21.
2.
Press
l
for current (upper display). If required,
press
l
again to activate loop power.
3.
Press
U
(lower display).
4.
If necessary, press
M
for SOURCE mode.
5.
Zero the pressure module.
6.
Perform checks at 0 % and 100 % of span and
adjust the transmitter as necessary.
Note
When HART resistor mode is selected, the
250
resistor is turned on both mA channels.