Fluke 726 FE 726 Users Manual - Page 57
Calibrating a Pressure Transmitter, UM
View all Fluke 726 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 57 highlights
Calibrating a Pressure Transmitter The following steps explain calibrating a pressure transmitter. 1. Connect the Calibrator to the instrument under test as shown in Figure 21. 2. Press l for current (upper display). If required, press l again to activate loop power. 3. Press U (lower display). Multifunction Process Calibrator Calibrating a Pressure Transmitter M 4. If necessary, press for SOURCE mode. 5. Zero the pressure module. 6. Perform checks at 0 % and 100 % of span and adjust the transmitter as necessary. Note When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels. 47
Multifunction Process Calibrator
Calibrating a Pressure Transmitter
47
Calibrating a Pressure Transmitter
The following steps explain calibrating a pressure
transmitter.
1.
Connect the Calibrator to the instrument under test
as shown in Figure 21.
2.
Press
l
for current (upper display). If required,
press
l
again to activate loop power.
3.
Press
U
(lower display).
4.
If necessary, press
M
for SOURCE mode.
5.
Zero the pressure module.
6.
Perform checks at 0 % and 100 % of span and
adjust the transmitter as necessary.
Note
When HART resistor mode is selected, the
250
Ω
resistor is turned on both mA channels.