Konica Minolta Kodak i780 Image Processing Guide - Page 73

Calibrate, Image, Chain, Image Chain, Do not use the calibration target, Start

Page 73 highlights

Use the calibration target provided with your scanner. 1. Click Calibrate. The Calibrate dialog box will be displayed. 2. Select Image Chain or UDDS from the drop-down box. • If you select Image Chain, use the calibration target provided with your scanner. Be sure to use a good, clean calibration target. Place the calibration target in the input elevator face up and click OK. IMPORTANT:If the White Background Accessory is installed, it must be replaced with the black background strips, and the scanner must be restarted prior to calibration. • If you select UDDS, place a 20 lb. or 75 g/m2 bond sheet of paper in the input elevator. Do not use the calibration target and click OK. 3. Click Start. While calibration is being performed a status will be provided in the Status box. 4. Click Done when the calibration is successful. 4-12 A-61504 October 2007

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85

4-12
A-61504
October 2007
Use the calibration target provided with your scanner.
1. Click
Calibrate
. The Calibrate dialog box will be displayed.
2. Select
Image
Chain
or
UDDS
from the drop-down box.
If you select
Image Chain,
use the calibration target provided
with your scanner. Be sure to use a good, clean calibration target.
Place the calibration target in the input elevator face up and click
OK
.
IMPORTANT:If the White Background Accessory is installed, it
must be replaced with the black background strips,
and the scanner must be restarted prior to
calibration.
If you select
UDDS,
place a 20 lb. or 75 g/m
2
bond sheet of paper
in the input elevator.
Do not use the calibration target
and click
OK
.
3. Click
Start
. While calibration is being performed a status will be
provided in the Status box.
4. Click
Done
when the calibration is successful.