Konica Minolta Kodak Scan Station 710 Administration Guide - Page 96
Uploading firmware, Scanner>Upload Firmware, Uploading, Please wait
View all Konica Minolta Kodak Scan Station 710 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 96 highlights
Uploading firmware 1. Select the Scan Station(s) where you want to send new firmware. NOTE: To select more than one Scan Station, use the Ctrl or Shift key and click the mouse button. The selected Scan Stations will be highlighted. 2. Select Scanner>Upload Firmware. The Open dialog box will be displayed. 3. Select the file that you want to upload and click Open. NOTES: • Only a KSSUpdate.exe file can be uploaded to the Scan Station. • The message, Uploading, Please wait will be displayed while the file is being uploaded. • If the Scan Station firmware and the KSSUpdate firmware are at the same revision level, no action will be taken. • You can "roll-back" to an older version or update to a newer version of the Scan Station firmware. If an older version of the KSSUpdate.exe file is uploaded, the Scan Station will be rolled back to use the older version of the firmware. If a newer version is uploaded, the Scan Station will be updated with this newer firmware. • You can obtain update packages from your Service Representative. Update packages will always be a file called KSSUpdate.exe. 9-12 A-61796 December 2013