3. ACP operation by using the LG ACCS 4 Enter ventilator information You can enter the ventilator group information from the 'System setting' menu. When you click on [Enter ventilator group] button located on the right side of the screen, ventilator group input window will be displayed. First, to add a new ventilator group, you must click on the applicable input block. When the input window switches to yellow color, enter the information of each item. Editing existing ventilators is applied in the same way. Click on the information block of the ventilator to edit and when it changes to yellow color, edit the existing information. To delete the existing ventilator information, click on the X button located on the left side. The applicable information will be deleted. Caution: Delete ventilator When you delete the ventilator group information, the ventilator information included in the applicable ventilator group will all be deleted. Information to center to set the ventilator group is as follows. Item Group number Name Description Enter the ventilator group number. The ventilator group number must not be duplicated and must be set within the range of 0~7. When you press the [Enter default value] button, you can easily create the number by receiving the higher level information. Enter the name of the ventilator group that the user can easily recognize. (Example. 1 floor) Installation/Owner's Manual 3-73