Panasonic AG-MDC20GJ / AG-MDR25PJ AG-MDR25 Camera Recorder Operating Instructi - Page 7

NOTICE FOR MEDICAL USE, Operating precaution

Page 7 highlights

NOTICE FOR MEDICAL USE ≥ This is an apparatus intended for recording playing-back of image from specified camera classified as; • Protection against Electric Shock Class II • Protection against Ingress of Water IPX0 • Mode of Operation Continuous • Not suitable for use in the presence of a Flammable Anaesthetic mixture with Air or with Oxygen or Nitrous Oxide. ≥ Equipment connected to signal input and/or output parts must be certified according to the appropriate IEC 60601-1 and/or IEC 60601-1 harmonized national standard. Furthermore all configurations shall comply with standard IEC 60601-1 ME system. Everybody who connects additional equipment to the signal input part, or signal output part configures a medical system, and is therefore responsible that the system complies with the requirements of the standard IEC 60601-1. If in doubt, consult the technical service department or your local representative. Leakage current of this unit may be exceeded the allowable value when conductively connected to other equipment. To avoid increment of the leakage current, separation device shall be applied. ≥ Use with the Compact Camera Head (AG-MDC20G: Optional). ≥ Do not use the Compact Camera Head in handheld style. ≥ Operator: To reduce the risk of electric shock, do not touch the patients during operating the equipment. ∫ TRANSPORT AND STORAGE CONDITION AMBIENT TEMPERATURE -10°C to 60°C RELATIVE HUMIDITY 30% to 80% ATMOSPHERIC PRESSURE 500 hPa to 1060 hPa Operating precaution Operating near any appliance which generates strong magnetic fields may give rise to noise in the video and audio signals. If this should be the case, deal with the situation by, for instance, moving the source of the magnetic fields away from the unit before operation. indicates safety information. - 7 -

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180
  • 181
  • 182
  • 183
  • 184
  • 185
  • 186
  • 187
  • 188
  • 189
  • 190
  • 191
  • 192
  • 193
  • 194
  • 195
  • 196
  • 197
  • 198
  • 199
  • 200
  • 201
  • 202
  • 203
  • 204
  • 205
  • 206
  • 207
  • 208
  • 209
  • 210
  • 211
  • 212
  • 213
  • 214
  • 215
  • 216
  • 217
  • 218
  • 219
  • 220
  • 221
  • 222
  • 223
  • 224
  • 225
  • 226
  • 227
  • 228
  • 229
  • 230
  • 231
  • 232
  • 233
  • 234
  • 235
  • 236
  • 237
  • 238
  • 239
  • 240
  • 241
  • 242
  • 243
  • 244
  • 245
  • 246
  • 247
  • 248
  • 249
  • 250
  • 251
  • 252
  • 253
  • 254
  • 255
  • 256
  • 257
  • 258
  • 259
  • 260
  • 261

- 7 -
indicates safety information.
NOTICE FOR MEDICAL USE
This is an apparatus intended for recording playing-back of image from specified camera classified as;
• Protection against Electric Shock
Class II
• Protection against Ingress of Water
IPX0
• Mode of Operation
Continuous
• Not suitable for use in the presence of a Flammable Anaesthetic mixture with Air or with Oxygen or Nitrous
Oxide.
Equipment connected to signal input and/or output parts must be certified according to the appropriate IEC
60601-1 and/or IEC 60601-1 harmonized national standard. Furthermore all configurations shall comply with
standard IEC 60601-1 ME system. Everybody who connects additional equipment to the signal input part, or
signal output part configures a medical system, and is therefore responsible that the system complies with the
requirements of the standard IEC 60601-1. If in doubt, consult the technical service department or your local
representative.
Leakage current of this unit may be exceeded the allowable value when conductively connected to other
equipment. To avoid increment of the leakage current, separation device shall be applied.
Use with the Compact Camera Head (AG-MDC20G: Optional).
Do not use the Compact Camera Head in handheld style.
Operator: To reduce the risk of electric shock, do not touch the patients during operating the equipment.
TRANSPORT AND STORAGE CONDITION
AMBIENT TEMPERATURE
-10°C to 60°C
RELATIVE HUMIDITY
30% to 80%
ATMOSPHERIC PRESSURE
500 hPa to 1060 hPa
Operating precaution
Operating near any appliance which
generates strong magnetic fields may
give rise to noise in the video and audio
signals.
If this should be the case, deal with the
situation by, for instance, moving the
source of the magnetic fields away from
the unit before operation.