Fluke 125 Fluke 125 Users Manual - Page 39
Triggering on a Waveform, Setting Trigger Level and Slope
View all Fluke 125 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 39 highlights
Triggering on a Waveform Triggering tells the test tool when to begin displaying the waveform. You can select which input signal should be used, on which edge this should occur, and you can define the condition for a new update of the waveform. You can setup the test tool to trigger on video signals. The bottom line of the waveform area identifies the trigger parameters being used. Trigger icons on the screen indicate the trigger level and slope. (See Figure 2-15.) Note: For 3 phase power measurements the trigger settings are fixed. 2 Scope/Meter Mode Triggering on a Waveform Setting Trigger Level and Slope For quick operation, use the AUTO SET key to automatically trigger on nearly all signals. c Perform an AUTO SET. To optimize trigger level and slope manually, do the following: c Press until you have left any open menu. d Enable the arrow keys for Trigger Level and Slope adjustment. e Adjust the Trigger Level continuously. Observe the trigger icon on the second time division line indicates the trigger level. f Trigger on either positive Slope or negative Slope of the chosen waveform. 2-17