Fluke 125 Fluke 125 Users Manual - Page 97
Horizontal, Trigger, Sensitivity A and B
View all Fluke 125 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 97 highlights
Horizontal Scope Modes Normal, Single, Roll Ranges Normal: equivalent sampling 10 ns to 500 ns/div real time sampling 1 μs to 5 s/div Single (real time 1 μs to 5 s/div Roll (real time 1s to 60 s/div Sampling Rate (for both channels simultaneously) Equivalent sampling (repetitive signals up to 1.25 GS/s Real time sampling: 1 μs to 5 ms/div 25 MS/s 10 ms to 60 s/div 5 MS/s Time Base Accuracy Equivalent sampling 0.4% +0.04 time/div) Real time sampling 0.1% +0.04 time/div) Glitch Detection 40 ns @ 20 ns to 5 ms/div ≥200 ns @ 10 ms to 60 s/div Glitch detection is always active. Horizontal Move 10 divisions Trigger point can be positioned anywhere across the screen. 10 Specifications Dual Input Oscilloscope Trigger Screen Update Free Run, On Trigger Source A, B, EXT EXTernal via optically isolated trigger probe ITP120 (optional accessory) Sensitivity A and B @ DC to 5 MHz 0.5 divisions or 5 mV @ 40 MHz 1.5 divisions @ 60 MHz 4 divisions Slope Positive, Negative Video on A interlaced video signals only Modes Lines, Line Select Standards PAL, NTSC, PAL+, SECAM Polarity Positive, Negative Sensitivity 0.6 divisions sync. 10-3