Fluke 125 Fluke 125 Users Manual - Page 82
Open the Input A or the Input B, Open the PROBE AC ADJUST
View all Fluke 125 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 82 highlights
Fluke 125 Users Manual c Open the application mode menu. k d Highlight SCOPE/METER e Enter the SCOPE/METER mode. f (A) Open the Input A or the Input B MEASUREMENTS menu. The F1....F4 button bar pops up. (B) g Open the INPUT... menu h Highlight AC ADJUST... i Open the PROBE AC ADJUST submenu. l A square wave appears on the screen. Adjust the trimmer screw in the probe housing to give an optimum square wave. Return to normal mode. j Highlight ADJUST 10:1 PROBE. 8-6
Fluke 125
Users Manual
8-6
Open the application mode menu.
Highlight SCOPE/METER
Enter the SCOPE/METER mode.
(A)
(B)
Open the Input A or the Input B
MEASUREMENTS menu. The
F1….F4 button bar pops up.
Open the INPUT… menu
Highlight
AC ADJUST…
Open the PROBE AC ADJUST
submenu.
Highlight ADJUST 10:1 PROBE.
A square wave appears on the
screen.
Adjust
the trimmer screw in the
probe housing to give an optimum
square wave.
Return to normal mode.