Fluke 125 Fluke 125 Users Manual - Page 92

Using Proper Grounding, Grounding with Short Ground Lead

Page 92 highlights

Fluke 125 User Manual Using Proper Grounding Incorrect grounding can cause various problems. This Section gives you guidelines for proper grounding. z Use the short ground lead(s) when measuring DC or AC signals on input A and input B. (See Figure 9-2.) Warning To avoid electrical shock or fire, use only one COM (common) connection , or ensure that all connections to COM are at the same potential. z Use the unshielded black ground lead to COM (common) for Ohm (Ω), Continuity, Diode, and Capacitance measurements. (See Figure 9-3.) Using the unshielded ground lead is also possible for single or dual input measurements for waveforms with a frequency up to 1 MHz. This may add some hum or noise to the waveform display due to the unshielded ground lead. Figure 9-2. Grounding with Short Ground Lead Figure 9-3. Grounding with Unshielded Ground Lead 9-6

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113

Fluke 125
User Manual
9-6
Using Proper Grounding
Incorrect grounding can cause various problems. This
Section gives you guidelines for proper grounding.
Use the short ground lead(s) when measuring DC or
AC signals on input A and input B. (See Figure 9-2.)
Warning
To avoid electrical shock or fire, use only one
COM (common) connection
, or ensure that
all connections to COM
are at the same
potential.
Use the unshielded black ground lead to COM
(common) for Ohm (
Ω
), Continuity, Diode, and
Capacitance measurements. (See Figure 9-3.)
Using the unshielded ground lead is also possible for
single or dual input measurements for waveforms with a
frequency up to 1 MHz. This may add some hum or
noise to the waveform display due to the unshielded
ground lead.
Figure 9-2. Grounding with Short Ground Lead
Figure 9-3. Grounding with Unshielded Ground Lead