Fluke 125 Fluke 125 Users Manual - Page 56

Performing Fieldbus Measurements, User 1

Page 56 highlights

Fluke 125 Users Manual Performing Fieldbus Measurements To perform fieldbus measurements do the following: c Open the application mode menu. d Highlight BUSHEALTH. e Open the bus selection menu: 4-2 f Select the bus type. Select User 1 or User 2 if you want to create a customized set of limits to test other (non standard) bus systems. See page 4-8 for setting the limits. Default settings are ControlNet for User1 and Foundation Fieldbus H1 for User2. g Accept your choice. For bus types followed by 3 dots ... a new menu will open: Use to select the required item and press . You will see a screen like fig 4.2. h Select the appropriate input A and input B test leads or probe type. i Connect the inputs as indicated in figure 4-1. Use the test tool inputs as shown in table 4-1.

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113

Fluke 125
Users Manual
4-2
Performing Fieldbus Measurements
To perform fieldbus measurements do the following:
Open the application mode menu.
Highlight BUSHEALTH.
Open the bus selection menu:
Select the bus type.
Select
User 1
or
User 2
if you
want to create a customized set
of limits to test other (non
standard) bus systems.
See
page 4-8 for setting the limits.
Default settings are ControlNet for
User1 and
Foundation Fieldbus
H1 for User2.
Accept your choice.
For bus types followed by 3 dots
… a new menu will open:
Use
to select the
required item and press
.
You will see a screen like fig 4.2.
Select the appropriate input A and
input B test leads or probe type.
Connect the inputs as indicated in figure 4-1.
Use the test tool inputs as shown in table 4-1.