Page 81 highlights
Note Ensure that the battery pack is placed in the battery compartment as shown in Figure 8-2. 8. Reinstall the battery cover and secure the screw. Using and Adjusting 10:1 Scope Probes Note The 10:1 voltage probe VPS40 that is supplied with Fluke 125 is always adjusted correctly to the Test Tool and needs no further adjustment. You need to adjust other 10:1 scope probes for optimal response. Warning To avoid electrical shock use the BB120 Banana-to-BNC adapter (delivered with the test tool) to connect a 10:1 scope probe to the input of the test tool. To adjust probes, do the following: • Connect the 10:1 scope probe from the gray input B jack to the red input A jack. Use the red 4-mm banana adapter (delivered with the probe) and the banana-toBNC adapter (BB120). See Figure 8-3. 8 Maintaining the Test Tool Using and Adjusting 10:1 Scope Probes BB120 Figure 8-3. Adjusting Scope Probes 8-5